Altitude
Warming-up time
30 minutes
Dimensions
Dimensions
width x length x height
Weight (incl. battery)
Maximum load
Vibration tolerance
Shock tolerance
Maximum total load
Image acquisition time
Estimated product life (if regularly
serviced and maintained according
to Agfa instructions)
Conversion screen
Pixel size
Active pixel matrix
Effective pixel matrix
Detector type
Active area size
Effective area size
DX-D 40C, DX-D 40G | Technical Data | 79
max. 2000 m
approx. 384 x 460 x 15.5 mm
< 3.4 kg
100 kg on an area of 40 mm in
diameter
0.7 G
1.6 G
150 kg over the whole detector
surface
6.5 s
Up to five (5) years
The minimal total dose that the panel
shall absorb during its lifetime is 74
Gy at RQA5
DR 40C
CsI:TI
140 µm
2560 x 3072
2548 x 3060
amorphous silicium
358 mm x 430 mm
356 mm x 428
mm
DR 40G
Gadox:Tb
2560 x 3072
358 mm x 430
mm
0290A EN 20140626 1622