Wireless Communication

Wireless communication is established between the internal wireless module
of the DR Detector and the NX workstation via the wireless access point. The
DR Detector is compliant with IEEE 802.11n (2.4 GHz/5 GHz). The available
frequency band varies depending on local radio laws and system
requirements. The frequency band (channel) of the DR Detector is selected at
installation.
DX-D 40C, DX-D 40G | Introduction to the DR Detector | 25
Note: Use of multiple pieces of equipment that use the same
frequency band (channel) may interfere with each wireless
communication and cause a decline in transmission speed.
Note: Before introducing other wireless equipment to the same
environment where the DR Detector is set up, consult the system
engineer or qualified personnel at the medical site.
Note: Do not place obstacles in the way of the wireless access point.
Otherwise, the properties of wireless communication, such as the
throughput and operable distance, may decrease.
Note: Transmitting the image data to the NX workstation takes a
number of seconds. After making an exposure, stay with the
detector in the direct neighbourhood (maximum 8 m) of the
wireless access point until the image is available on the NX
workstation.