Attaching the Zero Calibration Box
(When using the benchtop base)
WARNING
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes .
Be careful not to pinch hands in parts of the instrument or accessories
that open and close.Doing so may result in injury.
The zero calibration box is used to perform zero calibration for measurement of reflectance.
c Before performing zero calibration, attach the target mask for calibration.

Procedure
1
. Pull the sample holder toward you and keep it
open.
2
. Fit the projections of the zero calibration
box into the grooves on the CF-300 and
then close the sample holder to hold the
box in place.

 Notes on Use of Zero Calibration Box


c Take care not to scratch, touch, or make the inside of the zero calibration box dirty with such as fingerprints.

c If the inside of the zero calibration box gets dirty, wipe it with a soft, clean, dry cloth.

c If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the

zero calibration box. Then wipe off the liquid with a cloth dampened with water, and leave the box to dry.
c Should the inside of the zero calibration box get so dirty that it

cannot be cleaned, replace the box with a new one.