68 | DX-D 100 | Getting started

Stopping DX-D 100

To stop DX-D 100:
1. Stop NX and shut down the PC from the Windows Start menu or login
window.
2. Put the unit in parking position.
a) Use the tube-collimator handles to restore the default position of the
b) Press and hold the brake control while using the handles to fully
c) Lower the arm and fully insert the parking detent into the catch.
3. Lock the device.
• To lock the device using the key, switch the key on the control panel to
• To lock the device using the on/off keypad, press and hold the power
Two minutes remain before shutdown. The X-ray generator and the DR
Detector are powered off.
4. In a configuration with wireless DR Detector, power off the DR Detector:
a) turn off the DR Detector,
0187K EN 20191024 1609
collimator and the X-ray tube.
retract the telescopic arm and turn the column until the parking detent
is aligned with the catch.
WARNING:
Lowering the arm when the telescopic arm is not fully
retracted, can trigger the handle bar and cause unintended
displacement of the equipment. Fully retract the telescopic
arm before lowering the arm.
the "OFF" position.
1
"OFF" position
1.
button until the green status indicator light goes out.
Power button
1.
Enter button
2.
1
2