Installation thermal conductivity detector on a 6890 gas chromatograph
Table of Contents

Advertisement

Front detector
Auxiliary
Left most screw slot
Back detector
Figure 3.
Top rear panel cutouts
7.
Place the top rear panel on its left-most mounting screw. Use the screw
as a hinge and angle the panel while sliding each manifold ID tag
through its cutout in the panel, working from left to right. When all the
tags are through the panel, finish installing the panel on the GC.
8.
Install the RFI cover, the pneumatics cover, and the detector top cover.
9.
Connect the source gas lines to the manifold. See Figure 4.
Figure 4.
Gas line connections
10. Restore gas pressures and leak check all fittings.
Installing the flow manifold
Insert tip of
pliers here
Back inlet
Front inlet
9