Agilent Technologies G1563A Installation Manual page 12

Installation thermal conductivity detector on a 6890 gas chromatograph
Table of Contents

Advertisement

Installing the makeup and reference regulators (nonEPC detectors)
Installing the makeup and reference regulators
(nonEPC detectors)
1.
2.
3.
10
Install a screw in the front or back position of the pneumatics carrier.
Do not tighten the screw.
Install a screw in the regulator bracket. Do not tighten the screw.
Back
Front
Place the regulators in the front or back position of the pneumatics
carrier. Tighten the screws with a T-20 Torx screwdriver.
Connect the tube on the makeup gas regulator to the fitting labeled MU
on the flow manifold. Tighten the nut fingertight, then use the 7/16-inch
wrench to tighten it 3/4 of a turn. Follow the same procedure for
connecting the reference gas regulator to the fitting labeled Reference
on the flow manifold.