Positioning And Securing The Detector - Agilent Technologies G1563A Installation Manual

Installation thermal conductivity detector on a 6890 gas chromatograph
Table of Contents

Advertisement

Positioning and securing the detector

Positioning and securing the detector
1.
2.
Caution
Be careful to remove only the insulation within the scribed circle.
3.
12
Remove the round metal cutout on the oven top and the square plastic
cutout in the electronics carrier in the front or back detector position, if
necessary. Cut the metal circle with diagonal cutters so that the nibs are
connected to the piece removed. Cut the plastic in six places with
diagonal cutters. Discard the cutouts.
Lift out the die-cut insulation plug from the front or back detector
position, if necessary.
Carefully remove the scribed circle of insulation from the oven top to
create an opening into the oven.
Back
Front