Agilent Technologies G1563A Installation Manual page 18

Installation thermal conductivity detector on a 6890 gas chromatograph
Table of Contents

Advertisement

Connecting the detector
2.
∆ PRT leads
(thin, with pins)
Connect Filament
If the ∆ PRT is not properly connected to the detector interface card, filament
Caution
burnout can occur.
3.
16
Attach the filament leads to the TCD board in the two left-most
positions of the connector block. Insert a small flat-blade screwdriver
in the rectangular slot and depress it while you insert the filament lead
in the round hole. Polarity does not matter.
Similarly, connect the ∆ PRT leads to the two right-most positions in the
connector block. Polarity does not matter.
Filament leads
(thick, black)
Connect ∆ PRT
Connector block
TCD board