AGFA DX-G User Manual page 83

Hide thumbs Also See for DX-G:
Table of Contents

Advertisement

Room temperature
Environmental conditions for mobile installation (during operation)
In line with IEC721-3-3: 3M1 and 3K2 with following restrictions:
Room temperature
Relative humidity
Vibration
Barometric pressure/altitude
Highest pressure
Lowest pressure
Warming-up time
Cold start
Physical emissions
Noise emission (sound power level according to ISO7779)
During scanning
Standby
Heat emission per hour (typical values)
During continuous operation
(with standard erasure time)
Standby
RFID reader
Frequency
Bandwidth
Maximum power
Protocol
Plates and cassettes
Compatible cassette/IP types
DX-G
DX-G, DX-M | Technical Data | 83
between 10 °C and 40 °C
Agfa CR HM5.x Mammo: between
+20°Cand+30°C (68 - 86°F)
Other plates and cassettes: between
+15°Cand +30°C (59 - 86°F)
between 15%and75% (non-condensing)
40 - 200 Hz; 1 m/s²; sinusoidal vibration
106 kPa (at sea level)
70 kPa (3000 m above sea level)
3min.
max. 65dB(A)
max.45dB(A)
245 Wh / 836 BTU
92Wh / 314BTU
13.56 MHz
14 kHz
290 pW
MIFARE
Agfa CR HD5.x General, FLFS, AEC, Ex-
tremities.
2321G EN 20181126 1428

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dx-m5170/1005170/200

Table of Contents