Make sure to insert the cassette with the black (tube) side to the front and
with the shutter opening mechanism and the locking mechanism down,
towards the cassette slot.
2
Tube side must be faced to the operator.
1.
The shutter opening mechanism and locking mechanism must be
2.
pointed down, to the slot of the digitizer.
Figure 6: Cassette orientation
As a result, the digitizer starts erasing the image plate and the status
indicator is blue flashing.
When the digitizer has finished erasing the cassette, the status indicator is
constantly lit in blue.
4. Remove the cassette with the erased image plates from the cassette output
buffer.
5. Press the Erase button again to switch to normal mode.
One minute after the last cassette has been processed, the digitizer
switches from "erasure" mode to "normal mode".
Related Links
‑ G/DX-M Features
DX
Note:
To re-erase an image plate, you must push the Erase button at the
front side before you insert the cassette. After that, you have 10
seconds to enter a cassette. If you do not, the digitizer returns to the
standby mode.