Initialization mode selection

Initialization mode selection

The following initialization (init) modes are selectable:
Table 26
Initialization Mode Switches
1
Requires firmware B.06.40 or above. Modules without LAN on board, see G1369C LAN Interface Card
Using Stored
When initialization mode Using Stored is selected, the parameters are taken from
the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured previously
by one of the described methods.
Figure 29
1290 Infinity II High-Speed Pumps User Manual
SW 6
OFF
OFF
ON
Non-Volatile
RAM
Stored
Parameter
Using Stored (Principle)
SW 7
SW 8
Init Mode
ON
OFF
Using Stored
ON
ON
Using Default
OFF
OFF
1
DHCP
Active
Parameter
267