Table 5–1: Valve Option Operating States
MODE
VALVE
Sample/Cal
SAMPLE
Zero/Span
Shutoff
Sample/Cal
ZERO CAL
Zero/Span
Shutoff
Sample/Cal
SPAN CAL
Zero/Span
Shutoff
Sample/Cal
SAMPLE
Zero/Span
Shutoff
Sample/Cal
ZERO CAL
Zero/Span
Shutoff
Sample/Cal
SPAN CAL
Zero/Span
Shutoff
Sample/Cal
SAMPLE
Zero/Span
Sample/Cal
ZERO CAL
Zero/Span
Sample/Cal
SPAN CAL
Zero/Span
Sample/Cal
SAMPLE
Zero/Span
Sample/Cal
ZERO CAL
Zero/Span
Sample/Cal
SPAN CAL
Zero/Span
Optional Hardware and Software
CONDITION
Open to SAMPLE inlet
Open to IZS inlet
Closed
Open to ZERO/SPAN Valve
Open to IZS inlet
Closed
Open to ZERO/SPAN Valve
Open to SHUTOFF Valve
Open to PRESSURE SPAN Inlet
Open to SAMPLE inlet
Open to ZERO AIR SCRUBBER
Closed
Open to ZERO/SPAN Valve
Open to ZERO AIR SCRUBBER
Closed
Open to ZERO/SPAN Valve
Open to SHUTOFF Valve
Open to PRESSURE SPAN Inlet
Open to SAMPLE inlet
Open to IZS inlet
Open to ZERO/SPAN Valve
Open to IZS inlet
Open to ZERO/SPAN Valve
Open to PRESSURE SPAN inlet
Open to SAMPLE inlet
Open to ZERO AIR SCRUBBER
Open to ZERO/SPAN Valve
Open to ZERO AIR SCRUBBER
Open to ZERO/SPAN Valve
Open to PRESSURE SPAN inlet
47