36
Exposure release with high voltage
NOTE
P
Checking the preparation limit time
•
Select manual/analog operation on the MOBILETT XP Digital control panel (A/D but-
ton).
•
Close the collimator and set 60 kV/10 mAs on the control panel.
•
The control display for "exposure circuit ready" on the control panel must light up green.
•
Set the exposure switch (S27) to preparation and check the "exposure ready" acoustic
signal.
•
Hold 'Preparation' and check the preparation limit time of approx. 20 seconds. ERR 25
appears on the display (preparation time without exposure).
•
Acknowledge the error with the "collimator light" key on the control panel.
Release an exposure
•
X
Release an exposure.
•
"Exposure circuit ready" on the control panel lights up within 15 sec maximum.
Manual exposure termination
The user must be able to cancel an exposure at any time.
•
Set a midrange kV value (e.g., 70 kV) and the highest possible mAs value.
•
X
Release an exposure and interrupt it immediately. "ERR 39" appears on the display and
a succession of brief signal tones (exposure cancelled) sounds. Acknowledge the error
message with the "collimator light" key on the control panel.
MOBILETT XP Digital
If the "DAP" and/or "Remote Control" options are installed, they
can be tested for proper function during the following tests involv-
ing release of X-ray radiation.
The radiation indicator on the control panel lights up during the exposure; at the
same time an acoustic signal sounds and the "exposure circuit ready" light on
the control panel goes out.
SPR8-230.814.30.03.02
11.05
Page 36 of 70
CS SD 24
Start-up
0
0
0
0
Siemens AG
Medical Solutions