Exposure release with high voltage

NOTE
P

Checking the preparation limit time

Select manual/analog operation on the MOBILETT XP Digital control panel (A/D but-
ton).
Close the collimator and set 60 kV/10 mAs on the control panel.
The control display for "exposure circuit ready" on the control panel must light up green.
Set the exposure switch (S27) to preparation and check the "exposure ready" acoustic
signal.
Hold 'Preparation' and check the preparation limit time of approx. 20 seconds. ERR 25
appears on the display (preparation time without exposure).
Acknowledge the error with the "collimator light" key on the control panel.

Release an exposure

X
Release an exposure.
"Exposure circuit ready" on the control panel lights up within 15 sec maximum.

Manual exposure termination

The user must be able to cancel an exposure at any time.
Set a midrange kV value (e.g., 70 kV) and the highest possible mAs value.
X
Release an exposure and interrupt it immediately. "ERR 39" appears on the display and
a succession of brief signal tones (exposure cancelled) sounds. Acknowledge the error
message with the "collimator light" key on the control panel.
MOBILETT XP Digital
If the "DAP" and/or "Remote Control" options are installed, they
can be tested for proper function during the following tests involv-
ing release of X-ray radiation.
The radiation indicator on the control panel lights up during the exposure; at the
same time an acoustic signal sounds and the "exposure circuit ready" light on
the control panel goes out.