Making the Initial Carrier Frequency Tolerance
Measurement

Purpose

The initial carrier frequency tolerance measurement is designed to verify the
accuracy of the transmitter's initial carrier frequency.

Measurement Method

This is measured by integrating over the frequency deviations of the packets first 4
bits (the preamble bits). The result is either a positive or negative number in Hz
indicating the frequency difference from the specified nominal carrier frequency.
This measurement requires the signal to be demodulated to measure the frequency
deviation of each symbol. After demodulation, the frequency offset of each of the
preamble bits is measured and averaged.

Making the Measurement

1. Press the
2. Press the
The factory default settings provide a Bluetooth™ compliant measurement. For
NOTE
special requirements, you may need to change some of the settings. Any changes
from the default values may result in invalid measurement data.
Press
parameters for the current measurement to their default settings.
• Mode setup and frequency/channel parameters.
• Measurement setup parameters.
Chapter 3

Making the Initial Carrier Frequency Tolerance Measurement

Measure
key.
Init Carrier Freq Tol
,
Meas Setup
More (1 of 2)
Use the
and
Mode Setup
parameters for all measurements made within the current mode. For further
information, refer to Chapter 1 of this document.
These are measurement specific parameters that are changed using the
measurement setup (
Meas Setup
settings at any time by pressing
. For further information on measurement setup parameters, refer to
Defaults
the ESA-E Series Spectrum Analyzers Bluetooth™ Measurement Personality
User's Guide.
Making Bluetooth™ Measurements
key.
,
Restore Meas Defaults
keys to change these
Frequency Channel
) menu. Parameters can be returned to default
Meas Setup
,
More (1 of 2)
at any time to return all
and
Restore Meas
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4402bEsa-e e4404bEsa-e e4405bEsa-e e4407b

Table of Contents