Point Delay And Sweep Delay; Making A Point Delay Measurement; Making A Sweep Delay Measurement; Delay Description - Agilent Technologies 4286A User Manual

Rf lcr meter
Table of Contents

Advertisement

Point Delay and Sweep Delay

Delay functions stabilize the measurements by delaying the measurement start for a speci ed
time after the stimulus is set.

Making a Point Delay Measurement

1. Press
.
4
Delay Time
5
2. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POINT DELAY TIME .
3. Enter the point delay time and press one of the unit keys such as

Making a Sweep Delay Measurement

1. Press
.
Delay Time
4
5
2. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP DELAY TIME .
3. Enter the sweep delay time and press one of the unit keys such as

Delay Description

Point Delay
The measurement start for each point is delayed for a user-speci ed time after the
stimulus is set at each point.
Sweep Delay
The sweep start is delayed for a user-speci ed delay time after the trigger.
5-2
T ypical Functions
Figure 5-1. Point Delay and Sweep Delay
.