ADM Flow Meter

Making a Flow Measurement

Making a Flow Measurement
CAUTION
NOTE
NOTE
20
Make sure that the flow to be measured is within the range of
the flow meter (750 mL/min). Over ranging the flow meter can
damage the transducer.
Always turn the flow meter on before connecting it to a
flow source.
1
When the unit is powered on, it defaults to volumetric flow
mode. (If you want to use the mass flow mode, briefly
press the Power/Mode button.)
2
Connect the flexible tubing to the gas source to
be measured.
3
Wait for the reading to stabilize (1 to 2 seconds), then
observe the reading.
Low flows (less than 1 mL/min) may take up to 5 seconds for
the reading to stabilize
Agilent CrossLab CS ADM Flow Meter Operation Manual