4-1-4. Connection of Finished Inspection Jig II and Startup of Lens Adjustment Program
Note 1: Confirm that "4-1-3. Introduction of the Lens Adjustment Program" has been completed before this procedure is executed.
Note 2: In case of performing Finished inspection, be sure to perform in the condition that the lens is laid down (do not perform in the
lens standing condition).
Equipment
• Personal Computer
• Finished Inspection Jig II (AC 100 V - 240 V)
Note:
The cables and so on are not supplied as accessories.
• AC Cable (Sold on the Market)
• RS232C Cable (D-Sub 9 Pins (Male with Securing Screws), D-Sub 9 Pins (Female with Securing Screws)) Straight Type (Sold on the
Market)
• USB to Serial Cable (D-Sub 9 Pins (Male)) (Sold on the Market)
• M-BOX-JIG (J-6082-722-A) (Or the Jig supplied with Finished Inspection Jig (J-6082-645-A))
• Lens Adjustment Program (RomCheckerSVerXXX.zip)
1. Connection
Connect equipment and checking lens as shown in Fig. 4-1-3.
Note:
Lens Adjustment Program cannot start up without the connection to the finished inspection Jig II.
D-sub 9 Pins: Male
AC Cable
RS232C Cable
2. Startup of Lens Adjustment Program
1)
Unzip the Lens Adjustment Program (RomCheckerSVerXXX.zip) and save it.
2)
Start up "RomCheckerSVerXXX.exe" from an arbitrary folder.
SAL35F14G (1.4/35G) (35mm F1.4G)
Finished Inspection Jig II
D-sub 9 Pins: Female
USB to Serial Cable
Fig. 4-1-3
Fig. 4-1-4
Checking lens
4-5