- The part where the cuff is wrapped is at a different height than the heart. (A
difference of 10 cm (4 inches) in height may cause a variation in the blood pressure
value of up to 7 or 8 mmHg.)
- Body movement or conversing during measurement.
- Cuff wrapped over thick clothing.
- Pressure on the arm due to a rolled up sleeve.
• In the case of a cuff for adults, the cuff should be wrapped to a tightness that allows
two fingers to be inserted in between the cuff and the arm.
• The accuracy of a flashing measurement value that is out of the measurement
range cannot be guaranteed. Always check the patient's condition before deciding
what steps to take.
• Do not use the cuff if it is damaged or has holes.
• Only an OMRON GS CUFF2 can be used with this device. The use of any other
cuff may result in incorrect measurement.
Disposal
As there is a risk of environmental pollution, follow your applicable national and local
legal regulations regarding disposal or recycling of this
equipment.
The main
constituents of each part are listed in the table "Materials List".
As there is a risk of infection, do not recycle patient attachments such as cuffs, but
dispose of them as instructed by your facility's procedures and applicable
regulations.
Consult also the "Correct Disposal of This Product (Waste Electrical & Electronic
Equipment)" section of the Manual.

Transportation and storage conditions

Storage and transportation:
Temperature range: -20 to +60°C (-4 to 140°F)
Humidity range: 10 to 95%RH
Atmospheric pressure: 500 to 1060hPa
Warranty and lifetime
Warranty:
- blood pressure monitor: 3 years
- AC adapter: 1 years
- cuffs: this is considered a disposable item depending upon usage
Lifetime:
- blood pressure monitor: 5 years (only when appropriate inspection in performed)
- AC Adapter: 5 years
- tubes: no lifetime as these are considered disposable items