Omron JPN1 - Automatic Blood Pressure Monitor Manual

Omron JPN1 - Automatic Blood Pressure Monitor Manual

Also See for JPN1:

Advertisement

Omron JPN1 - Automatic Blood Pressure Monitor Manual

Overview

Overview - Main unit:

  1. Display
  2. START/STOP button
  3. Memory button
  4. Up/Down () buttons
  5. Buzzer and Date/Time setting () button
  1. Air Jack
  2. Battery compartment
  3. AC adapter jack (for optional AC adapter)

Arm cuff:
Overview - Arm cuff

  1. Arm cuff (Medium cuff: arm circumference 22-32cm)
  2. Marker
  3. Air tube
  4. Air plug

Display:
Overview - Display

  1. Systolic blood pressure
  2. Diastolic blood pressure
  3. Heartbeat symbol (Flashes during measurement)
  4. Memory symbol
  5. Deflation symbol
  6. Average value symbol
  1. Movement error symbol
  2. Date/Time display
  3. Blood pressure level indicator
  4. Pulse display
  5. Battery low symbol
  6. Irregular heartbeat symbol

Preparation

Installing/Replacing the Batteries

Installing/Replacing the Batteries

  1. Remove the battery cover.
  2. Insert four "AA" batteries as indicated in the battery compartment and then replace the battery cover.

Notes:

  • If the battery low symbol () appears on the display, turn the unit off then replace all batteries at the same time.
  • The measurement values continue to be stored in memory even after the batteries are replaced.

Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.

Setting the Buzzer/ Date and Time

  1. Press the button.
  2. Set the monitor to the correct date and time before taking a measurement for the first time.
    Setting the Buzzer/ Date and Time
  3. Press the START/STOP button to store the setting.
    Notes:
  • If the batteries have been removed for 30 seconds or more, the date and time setting will need to be reset.
  • If the date and time are not set, "-:--" appears during or after measurement.

Using the Unit

Applying the Arm Cuff

Remove tight-fitting clothing or tight rolled up sleeve from your upper arm.
Do not place the cuff over thick clothes.

  1. Insert the air plug into the air jack.
    Applying the Arm Cuff
  2. Put your arm through the cuff loop.

  3. Position the arm correctly.

    The bottom edge of the cuff should be 1 to 2 cm above the elbow.
    Marker (arrow under tube) is centred on the middle of your inner arm. Close the fabric fastener FIRMLY.
    Notes:
  • When you take a measurement on the right arm, air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube.
  • The blood pressure can differ between the right arm and the left arm, and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your doctor which arm to use for your measurement.

How to Sit Correctly

To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No eating, smoking or exercising 30 minutes before taking a measurement.

  • Sit on a chair with your feet flat on the floor.
  • Sit upright with your back straight.
  • The cuff should be at the same level as your heart.

Taking a Reading

Notes:

  • To cancel a measurement, press the START/STOP button to turn off the unit and to release the air in the arm cuff.
  1. Press the START/STOP button. The cuff will start to inflate automatically.

    (If your systolic pressure is more than 220 mmHg
    After the cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
    Notes:
  • The monitor will not inflate above 299 mmHg.
  • Do not apply more pressure than necessary.)
    Note: Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.


Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.

  1. Remove the arm cuff.
  2. Press the START/STOP button to turn off the monitor. The monitor automatically stores the measurement in its memory.
    It will automatically turn off after five minutes.
  • Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home.
    Systolic Blood Pressure Above 135 mmHg
    Diastolic Blood Pressure Above 85 mmHg
    This criteria is for home blood pressure measurement.
  • Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats can influence the results of the measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the irregular heartbeat symbol ( ). If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat the measurement. If the irregular heartbeat symbol ( ) is shown frequently, please make your doctor aware of it.

Using the Memory Function

The monitor automatically stores the result up to 60 sets. It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average will be based on two readings. If there is one reading in memory for that period, the average will be based on one reading.
Notes:

  • If the memory is full, the monitor will delete the oldest readings.
  • When viewing the reading taken without setting the date and time, "-:--" is displayed instead of the date and time.

Using the Memory Function

View the Average Value

  • If the previous reading was taken without setting the date and time, the average value is not calculated.

View the Previous Reading

  • The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered "1".

Note: You cannot partially delete the stored readings.

To delete all the values stored in memory
When the memory symbol ( ) appears, first press the Memory button. Then while holding it down, press the START/STOP button simultaneously for about 2-3 seconds.
To delete all the values stored in memory
Note: You cannot partially delete the stored readings.

Troubleshooting and Maintenance

The Icons and Error Messages

Error Display Cause Remedy
Irregular heartbeats are detected. Remove the arm cuff. Wait 2-3 minutes and then take another measurement. If this error continues to appear, contact your doctor.
Movement during measurement. Carefully read and repeat the steps.
Blink The batteries are low. You should replace them with new ones ahead of time.
Lit The batteries are exhausted. You should replace them with new ones at once.
Cuff is under inflated. Carefully read and repeat the steps.
Movement during measurement. Repeat measurement. Remain still and do not talk during measurement.
Air plug disconnected. Insert the air plug securely.
Arm cuff not applied correctly. Apply the arm cuff correctly.
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff.
Air is leaking from the arm cuff. Replace cuff with new one.
The arm cuff was inflated above 299 mmHg when inflating the cuff manually. Do not inflate the arm cuff above 299 mmHg.
Device error. Contact your local OMRON representative.

Note: The irregular heartbeat symbol () may also be displayed with error messages.

Troubleshooting

Problem Cause Remedy

The reading is extremely low (or high)

Arm cuff not applied correctly. Apply the arm cuff correctly.
Movement or talking during measurement. Remain still and do not talk during measurement.
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff.

Arm cuff pressure does not rise

The air tube is not securely connected into the main unit. Make sure that the air tube is connected securely.
Air is leaking from the arm cuff. Replace the arm cuff with a new one.

Arm cuff deflates too soon

The arm cuff is loose. Apply the cuff correctly so that it is firmly wrapped around the arm.
Cannot measure or readings are too low or too high. The arm cuff has not been inflated sufficiently. Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result.

Nothing happens when you press the buttons

The batteries are empty. Replace the batteries with new ones.
The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity.

Other problems

  • Press the START/STOP button and repeat measurement.
  • If the problem continues, try replacing the batteries with new ones.
If this still does not solve the problem, contact your local OMRON representative.

Maintenance

To protect your unit from damage, please observe the following:

  • Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight.
  • Do not fold the cuff or tubing tightly.
  • Do not inflate the arm cuff over 299 mmHg.
  • Do not disassemble the unit.
  • Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).
  • Do not use volatile liquids to clean the main unit.
  • Do not wash the arm cuff or immerse it in water.
  • Do not use petrol, thinners or similar solvents to clean the arm cuff.
  • Do not carry out repairs of any kind yourself. If a defect occurs, consult your local OMRON representative.
    Maintenance
  • The unit should be cleaned with a soft, dry cloth.
  • Use a soft, moistened cloth and soap to clean the arm cuff.

Calibration and Service

  • The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.
  • It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy. Please consult your local OMRON representative.

Storage

  1. Unplug the air tube from the air connector.
  2. Gently fold the air tube into the arm cuff.
    Note: Do not bend the air tube excessively.

    Do not store the unit in the following situations:
  • If the unit is wet.
  • Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.
  • Locations exposed to vibrations, shocks or where it will be at an angle.

Optional Parts

Medium Arm Cuff
Arm circumference 22 - 32 cm
Large Arm Cuff
Arm circumference 32 - 42 cm
AC Adapter

Note: Please check with your local OMRON representatives for the appropriate optional part models.

Using the Optional AC Adapter
Using the Optional AC Adapter

  1. Insert the AC adapter plug into the AC adapter jack on the rear side of the main unit.
  2. Plug the AC adapter into an electrical outlet.
    To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the main unit.

Technical Data

Product Description Automatic Blood Pressure Monitor
Model JPN1
Display LCD Digital Display
Measurement Method Oscillometric method
Measurement Range Pressure: 0 to 299 mmHg Pulse: 40 to 180/min.
Accuracy Pressure: ±3 mmHg Pulse: ±5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Memory 60 Measurements with date and time
Rating DC6V 4W
Power Source 4 "AA" batteries 1.5V or AC adapter (optional, INPUT AC100-240V 50/60Hz 0.12A)
Battery Life Approx. 1500 measurements with new alkaline batteries at a room temperature of 23°C
Applied Part = Type B
Protection Against Electric Shock Internally powered ME equipment
(When using only the batteries)
= Class II ME equipment (Optional AC adapter)
Operating temperature/ Humidity +10 to +40°C / Maximum: 30 to 90% RH
Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH
Humidity/ Air pressure / 700-1060 hPa
Console Weight Approx. 340g without batteries
Cuff Weight Approx. 120g
Outer Dimensions Approx. 123 (w) mm × 85 (h) mm × 141(l) mm
Cuff Dimensions Approx. 146 mm × 446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Cuff Material Nylon and polyester
Package Content Main unit, cuff, instruction manual, battery set

Note: Subject to technical modification without prior notice.

Documents / Resources

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Omron JPN1 - Automatic Blood Pressure Monitor Manual

Advertisement

Need Assistance?

Do you have a question about the JPN1 that isn't answered in the manual? Leave your question here.

Error while submitting your question. Please, reload page and try again.

Thank you! Your question has been received!

Table of Contents