- The patient uses an artificial heart and lung (there will be no pulse)
- The patient has a mastectomy
- The patient has an aneurysm
- The patient has frequent arrhythmia
- Body motions such as convulsions, arterial pulsations, or trembling (cardiac
massage in progress, minute continuous vibrations, rheumatism, etc.)
• Before use, visually inspect the unit to make sure there are no deformations due to
falling, and that there is no dirt or moisture on the unit.
• When the unit has not been used for an extended period of time, always verify that
it operates normally and safely before use.
• Do not use in a location where the unit may easily fall. In the event that the unit
falls, verify that it operates normally and safely.
• Do not wrap the cuff on an arm with an SpO2 sensor or other monitoring equipment
attached. The pulse may disappear when the cuff pressurizes, causing a temporary
loss of monitoring function.
Cleaning
Warnings
• When cleaning the unit, turn off the power and disconnect the AC adapter from the
unit.
• After cleaning the unit, make sure it is completely dry before connecting to a power
outlet.
• Do not spray, pour, or spill liquids into or onto the unit, accessories, connectors,
buttons, or openings in the housing.
Cautions
• Do not use thinner, benzene, or other solvents to clean the unit.
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-
concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the
manufacturer. This may damage the surface of the device.
• Clean the unit regularly
.
Maintenance and inspection
Warnings
• To use the unit safely and correctly, always inspect the unit when starting work.
• Unauthorized modification is prohibited by law. Do not attempt to disassemble or
modify the unit.
Dry cell battery
Warnings
• If battery fluid comes in contact with the eye, immediately flush with copious
amounts of water. Do not rub. Seek medical attention immediately.
• Do not throw into flame, disassemble, or heat.
• Always disconnect the AC adapter from the unit before removing or
installing a battery.