6.5.5.4

Quick Start Wizard via the handheld programmer

Note
• A reset to factory defaults should be performed before running the Quick Start Wizard if
the device has been used in a previous application. See Quick Start Wizard via the local
display push buttons (Page 45).
• The Quick Start wizard settings are inter-related and changes apply only after you select
Finish in the Wizard Complete step.
• Do not use the Quick Start wizard to modify parameters: see instead Parameter
Reference (Page 117). Perform customization for your application only after the Quick
Start has been completed.
Quick Start Wizard
1. Point the programmer at the display from a maximum distance of 300 mm (1 ft.), then
press RIGHT
2. Press RIGHT
3. Press RIGHT
directly to the next item.
4. To change a setting, scroll to the desired item or key in a new value.
5. After modifying a value, press RIGHT
the next item.
6. At any time, you can press UP
Measurement mode.
LR560 (FOUNDATION FIELDBUS)
Operating Instructions, 06/2018, A5E34648692-AE
to activate PROGRAM mode and open menu level 1.
twice to navigate to menu item 1.1 and open parameter view.
to open Edit mode or DOWN
to go back, or LEFT
6.5 Commissioning via handheld programmer
to accept default values and move
to accept it and press DOWN
to cancel and return to
Commissioning
to move to
53