4.2.5
Aimer adjustment
Note
Measuring material in the cone
Aiming will assist in measuring material in the cone.
①
3" flange
①
②
1. For 4 and 6" Aimer: loosen the set screws in the locking ring.
– Holding the electronics enclosure firmly, loosen the Aimer locking ring using the
2. Direct SITRANS LR560 so the antenna is pointed at an angle perpendicular to the
material surface, if possible.
3. When the desired position is reached, re-tighten the locking ring using the C spanner, and
tighten set screws.
LR560 (FOUNDATION FIELDBUS)
Operating Instructions, 06/2018, A5E34648692-AE
Aiming is not required for signal optimization with 78 GHz frequency.
Aiming will assist in measuring material in the cone.
supplied C spanner, until the LR560 drops down slightly. The enclosure can then be
turned freely.
②
4" and 6" flange
Installing/mounting
4.2 Installation instructions