Lcd Contrast (4.10.); Secondary Value (4.11.); Simulate Enable (4.12.); Demo Mode (4.13.) - Siemens SITRANS LR560 Operating Instructions Manual

Radar transmitters (foundation fieldbus)
Hide thumbs Also See for SITRANS LR560:
Table of Contents

Advertisement

8.4.10

LCD CONTRAST (4.10.)

Factory setting:
Setting range:
Purpose:
Description:
8.4.11

SECONDARY VALUE (4.11.)

The value displayed in the secondary region of the LCD, in Measurement Mode, see The
LCD display (Page 43), area (6) under Normal operation.
Use Secondary Value to capture the menu navigation path to a selected parameter, and
store a custom secondary value, for example ECHO STRENGTH (2.4.7.2.) (Page 132).
While in Parameter View
stores the path to the selected parameter in Secondary Value, and displays that value in the
secondary region of the LCD display when in Measurement Mode.
See Parameter view under The LCD display (Page 43).
1)
8.4.12

SIMULATE ENABLE (4.12.)

Replaces a physical jumper switch found on some FF devices to enable simulation when set
to ON. (Available only via local operation.)
Options
For more information on Simulation, see Simulation (input) under Operation (LTB) (Page 74)
in AMS Device Manager, or the manual,
(7ML19985MP01)
8.4.13

DEMO MODE (4.13.)

Factory setting:
Setting range:
Purpose:
LR560 (FOUNDATION FIELDBUS)
Operating Instructions, 06/2018, A5E34648692-AE
8
0 to 20
The factory setting is for optimum visibility at room temperature and in av-
erage light conditions. Extremes of temperature will lessen the contrast.
Contrast setting will depend on ambient temperature.
Adjust the value to improve visibility in different temperatures and light
conditions. Change the value in small steps to ensure you can continue to
read the display.
mode of the selected parameter, press the decimal point key. This
1)
*
OFF
Simulation Disabled
ON
Simulation Enabled
Foundation Fieldbus for Level instruments
.
OFF
ON or OFF
Used to set up for demonstrations: reduces the time between measure-
ments and the accuracy for demonstration purposes.
Parameter Reference