4 = Four Calibration Points, Cont.
4. With the "Known test weight" screen displayed, place the first calibrated test weight
(one fourth the test load) on the scale, enter the setting for the test weight and then
press the Save key.
CALIBRATE SCALE
250.0
This is the known
test weight
that is currently
on the scale.
SAVE
CANCEL
5. The display will change to show "Working" and after a few seconds change to show the
"Known test weight" screen.
6. With the "Known test weight" screen displayed, place the second calibrated test weight
(half the test load) on the scale, enter the setting for the test weight and then press the
Save key.
CALIBRATE SCALE
500.0
This is the known
test weight
that is currently
on the scale.
SAVE
CANCEL
7. The display will change to show "Working" for a few seconds.