The Sonar Height Rod on your icon Digital Clinical Scale has been pre-configured and
calibrated at the factory and should not require changes. However, if the factory settings do not
meet the requirements of your operation, the following describes the process to calibrate the
sonar height rod, change the height setting, or disable the Sonar Height Rod.
To Enter Sonar Calibration
1. Press the
key to turn the scale on.
2. The scale will show the Detecto logo briefly and then change to the Weight screen.
3. With the Weight screen displayed, press the Menu key.
4. The display will change to show the Settings Overview 1 screen.
SETTINGS OVERVIEW 1
Revision 1.00.09, Display Rev. 1.0.18
Audit Trail -- Cal=4, Cfg=2
Range 1: 600.0 x 0.2 -- 2: 1000.0 x 0.5
Weight Units=lb
Auto Lock=5 s
Motion=3 div.
OIML=No
USB: SMA
Power up zero=No
Sonar Exit Setup
5. Touch the Sonar key.
6. The display will change to a selection screen to disable or calibrate the sonar height rod,
or to set a new height from the scale platform to the bottom of the sonar sensor.
CALIBRATE SONAR
0.0
0 = Disable
1 = Calibrate
Or enter a new
HEIGHT in cm.
SAVE
CANCEL
To Disable the Sonar Height Rod
1. Press the 0 key and then press the Save key.
2. The display will change to show the Settings Overview 1 screen.
3. The sonar height rod is now disabled.
4. Press the Exit key to return to the Weight screen.
28
3300-0192-0M Rev H icon Clinical Scale Owner's

SONAR CALIBRATION

>>>
Height Units=in
Sonar Ht=221.5 cm
Filter=Off
ZeroTrack=10/2 d
Gravity=0.000000
1
2
3
4
5
6
7
8
9
0
.
DEL