Detecto ICON Series Owner's Manual page 35

Digital clinical scale with wi-fi and bluetooth 5.0
Table of Contents

Advertisement

To Calibrate the Sonar Height Rod
1. Press the 1 key and then press the Save key.
2. The display will show "Clear" for approximately four seconds indicating you should step
away from the sensor at a distance of at least 6 feet as it prepares to calibrate.
NOTE: Movable objects that might interfere with the sound waves of the sensor should
be removed from a 6-foot radius. Objects that are not easily movable (walls, a desk,
large heavy objects that stay in one-spot) can be left in place. These larger objects will
be masked away in the calibration process.
CALIBRATE SONAR
Clear
0 = Disable
1 = Calibrate
Or enter a new
HEIGHT in cm.
SAVE
CANCEL
3. After approximately four seconds, the display will briefly show "Working" as the sonar
sensor calibrates. At this time, the sensor is recording data on its surroundings and auto-
detecting the distance to the scale platform.
NOTE: Calibration will take approximately ten seconds.
CALIBRATE SONAR
Working
0 = Disable
1 = Calibrate
Or enter a new
HEIGHT in cm.
SAVE
CANCEL
4. When calibration is complete, the scale will beep, and then the display will return to the
Settings Overview 1 screen.
5. The sonar height rod has now been calibrated to its surroundings.
6. Press the Exit key to return to the Weight screen.
IMPORTANT! Any time the scale is moved to a different location, it should
be recalibrated.
3300-0192-0M Rev H icon Clinical Scale Owner's