Detecto ICON Series Owner's Manual page 46

Digital clinical scale with wi-fi and bluetooth 5.0
Table of Contents

Advertisement

3 = Three Calibration Points, Cont.
4. With the "Known test weight" screen displayed, place the first calibrated test
weight (half the test load) on the scale, enter the value for the test weight and then
press the Save key.
CALIBRATE SCALE
500.0
This is the known
test weight
that is currently
on the scale.
SAVE
CANCEL
5. The display will change to show "Working" for a few seconds.
CALIBRATE SCALE
Working
This is the known
test weight
that is currently
on the scale.
SAVE
CANCEL
6. With the "Known test weight" screen displayed, place the second calibrated test
weight (full test load) on the scale, enter the setting for the test weight and then
press the Save key.
CALIBRATE SCALE
1000.0
This is the known
test weight
that is currently
on the scale.
SAVE
CANCEL
7. The display will change to show "Working" for a few seconds and then return to the
Settings Page 5 screen.
8. Calibration is now complete.
9. Press the Exit key to exit Setup and return to the Weight screen.
40
3300-0192-0M Rev H icon Clinical Scale Owner's
1
2
3
4
5
6
7
8
9
0
.
DEL
1
2
3
4
5
6
7
8
9
0
.
DEL
1
2
3
4
5
6
7
8
9
0
.
DEL
NOTE: A minimum of 50%
of the scale's capacity is
required for the full test
load, however 70% to 100%
is recommended.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

IconIcon-ukIcon-euIcon-lxiIcon-uwaIcon-eu-ver

Table of Contents