To Enter Calibration, Cont.
SETTINGS
Calibrate Scale
Load Cell Trim
Power up Zero: No
Auto Locking: 5 s
<<<
8. Select Calibrate Scale from the Settings Page 5 screen.
9. The display will change to show the "Number of Points" screen.
CALIBRATE SCALE
2
Number of points
at which the
scale will be
calibrated.
1 = SmartCal
SAVE
CANCEL
10. With the "Number of Points" screen displayed, select 1, 2, 3, or 4 on the touchscreen
and then press the Save key.
1 = SmartCal Calibration
(calibrated load placed over each load cell only once)
2 = Two Calibration Points
(empty scale and full test load)
3 = Three Calibration Points
(empty scale, half test load, and full test load)
4= Four Calibration Points
(empty scale, one fourth test load, half test load, and full test load)
11. Proceed to next sections of the manual for the type of calibration selected.