Detecto ICON Series Owner's Manual page 30

Digital clinical scale with wi-fi and bluetooth 5.0
Table of Contents

Advertisement

Settings Page 5
SETTINGS
Calibrate Scale
Load Cell Trim
Power up Zero: No
Auto Locking: 5 s
<<<
Calibrate Scale
Touching this key will allow for calibration of the scale. Refer to the CALIBRATION section
of this manual for instructions to select the calibration mode and how to perform calibration.
Load Cell Trim
Touching this key will allow for trimming (adjusting the output of) the load cells in the scale.
Refer to the LOAD CELL TRIM section of this manual for instructions on how to trim the
load cells.
Auto Locking:
This setting is used by the scale to hold a stable patient weight for a desired amount of time.
For example, if a setting of 5 seconds is used, then when the scale locks onto a stable
patient weight, it will remain locked for 5 seconds before automatically releasing the weight.
Touching this key will open an entry screen for the auto-locking feature of the scale. The
setting entered is a time in seconds. Using the numeric keys, enter a new setting, and then
press the Save key.
Allowable settings for Auto Locking are: 0 to 60.
Power up Zero:
Touching this key will toggle the power up zeroing of the scale to Yes or No. If enabled, this
will cause the scale to attempt to zero the scale on power up.
Allowable settings for Power up Zero are: Yes or No.
Press the >>> key to advance to the Settings Page 6.
Press the <<< key to return to the Settings Page 4.
Press the Exit key to exit Setup and return to the Weight screen.