To Enter A New Height
The height in centimeters from the scale platform to the bottom of the sonar height rod
sensor has been pre-configured at the factory and should not require changes. If an
additional mat or other covering has been added to the scale platform, the new distance,
in centimeters from the scale platform to the bottom of the sonar sensor must be measured
and entered as accurately as possible to guarantee accurate height measurements from the
sonar height rod.
CALIBRATE SONAR
0.0
0 = Disable
1 = Calibrate
Or enter a new
HEIGHT in cm.
SAVE
CANCEL
1. Using the numeric keys, enter the new height in centimeters, (including the decimal point
if needed).
2. Press the Save key.
3. The display will change to show the Settings Overview 1 screen.
4. Press the Exit key to return to the Weight screen.
IMPORTANT! If a new height has been entered (distance from scale platform
to bottom of the sonar sensor), the scale should be recalibrated.
30
3300-0192-0M Rev H icon Clinical Scale Owner's
1
2
3
4
5
6
7
8
9
0
.
DEL