1 = SmartCal, Cont.
8. The display will change to show "Working" for a few seconds and then change to show
"Set test weight over Cell 3".
CALIBRATE SCALE
Set test weight
over Cell 3 and
press Save.
SAVE
CANCEL
9. Place the calibrated test weight on the scale over Cell 3 and then press the Save key.
10. The display will change to show "Working" for a few seconds and then change to show
"Set test weight over Cell 4".
CALIBRATE SCALE
Set test weight
over Cell 4 and
press Save.
SAVE
CANCEL
11. Place the calibrated test weight on the scale over Cell 4 and then press the Save key.
12. The display will change to show "Working" for a few seconds and then return to
the Settings Page 5 screen.
13. Calibration is now complete.
14. Press the Exit key to exit Setup and return to the Weight screen.