WI-FI AND BLUETOOTH SMA COMMANDS, CONT.
Command
Hex Rep.
N – scale
0A4E0D
information
scroll
R – Repeat
0A520D
Displayed
Weight
Continuously
XB – battery level
0A58420D
percentage
62
3300-0192-0M Rev H icon Clinical Scale Owner's
Response
Each time sent you will get the next line of scale
information until there is no longer any information.
1. 0A 54 59 50 3A 53 0D
= TYP:S
2. 0A 43 41 50 3A 20 6C 62 3A 36 30 30 2E 30 3A 32 3A
31 0D
= CAP: lb:600.0:2:1, this depends on the settings of the
Icon.
600.0 – Capacity, 2 – Interval, & 1 – Decimal
3. 0A 43 4D 44 3A 48 52 49 4E 58 0D
= CMD:HRINX
4. 0A 45 4E 44 3A 0D
= END:
5. If N is sent again you will get the unknown command
response until the I command is sent again. 0A 3F 0D
= ?
0A 5A 31 47 20 20 30 30 30 30 30 30 2E 30 30 6C 62 0D
= Z1G 000000.00lb, you should get this continuously
until another SMA command is received.
0A 38 36 2E 32 35 0D
= 86.25