8
Test Functions
D/A Converter (DAC) Test
Test Evaluation
The noise on the step should be less than 3 µV.
Probable cause
1
2
Agilent InfinityLab LC Series RID User Manual
Bad cable or grounding problem between
detector and external device.
Defective detector main board.
Suggested actions
Check or replace the cable.
Please contact your Agilent service representa-
tive.