Test Evaluation - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

8
Test Functions
D/A Converter (DAC) Test

Test Evaluation

The noise on the step should be less than 3 µV.
Probable cause
1
2
Agilent InfinityLab LC Series RID User Manual
Bad cable or grounding problem between
detector and external device.
Defective detector main board.
Suggested actions
Check or replace the cable.
Please contact your Agilent service representa-
tive.