Programming and Operation

Rough Valve Open/Closed

Last
1
2
Next
4
5
Clear
7
8
-
0
3-48
Opening the rough valve when the On-Board Cryopump is turned ON
may cause oil from the roughing pump to backstream into the vacuum
chamber and contaminate the cryopump arrays. A message is dis-
played on the keypad/display warning you about this potential problem.
The fourth parameter of the CONTROL function allows you to open or
close the rough valve as shown in Figure 3-43.
ROUGH V OPEN
3
MONITOR
CONTROL
6
REGEN
RELAYS
9
SERVICE
HELP
Enter
oo
®
.
ON-BOARD
Figure 3-43: Rough Valve Open/Closed Display