On-Board FastRegen Control Module Programming and Operation Instructions

Extended Purge

Repurge

P/N 8040406
The fifth parameter of the REGEN function allows you to extend the purge
time during Full regeneration to ensure the cryopump has been purged of
all contaminants. The extended purge time screen display is shown in
Figure 3-16.
Last
3
1
2
Next
4
5
6
Clear
7
8
9
-
0
.
Figure 3-16: Extended Purge Display
The sixth parameter of the REGEN function allows you to establish the
repurge time in the event of repurge as shown in Figure 3-17. A repurge
occurs during Full regeneration when the roughing pressure fails to
decrease at a rate of at least 2% per minute.
Last
3
1
2
Next
4
5
6
Clear
7
8
9
-
0
.
Figure 3-17: Repurge Display
EXTD PURGE 10M
MONITOR
CONTROL
REGEN
RELAYS
SERVICE
HELP
Enter
®
oo
ON-BOARD
10 MIN REPURGE
MONITOR
CONTROL
REGEN
RELAYS
SERVICE
HELP
Enter
®
oo
ON-BOARD
-
C
TI
CRYOGENICS
HELIX TECHNOLOGY CORPORATION
3-15