6-5. Time Setup & Display
Sets time and display of the SRX-101A's internal clock. The time set here will be used to specify
time for each event in the log and displayed together with the events.
1. DIP Switch setting
2. Operation Method
1) Set the DIP switches as shown above, and turn on the [Power] breaker.
· "Year" of the time display mode is displayed by the 7seg.LED.
2) Time display mode
· Each time when the [Repl.] SW is pressed;
"Year" display → "Month" display → "Date" display → "Hour" display → "Minute"
display → "Second" display → "Year" display → repeats in order.
· Each time when the [RUN] SW is pressed;
"Year" display → "Second" display → "Minute" display → "Hour" display → "Date"
display → "Month" display → "Year" display → repeats in order.
3) Changing the time display mode to time setup mode
Pressing the [RUN] SW and [Repl..] button simultaneously will allow the display "c00" to
blink.
Keep pressing the two buttons for three seconds or more will change the display from
blinking to continuous display, and releasing your fingers off the buttons will enter the
"Time Setup" mode.
Releasing your fingers off the buttons within three seconds will not enter the "Time Setup"
mode, and it stays in the "Time Display" mode.
4) Time Setup Mode
· Change the set item using the [Repl.] SW;
"Year" data change → "Month" data change
change → "Minute" data change → "Second" data change → returns to "Time Display"
mode
· Displayed value can be changed using the [RUN] SW.
DS1-1
ON
DS2
[2]
CPU BOARD DIP SWITCHES
→ "Date" data change
→ "Hour" data