b. Quick Procedure
If you are in a hurry to fill the tanks, you can use the funnel provided to fill the tanks manually.
Begin the following procedures with only the DEV rack set in place.
(1)
Push the OFF side (O side) of the power breaker.
(2)
Fill the FIX tank with solution up to the full mark, begin careful not to splash any of the
solution into the DEV tank.
NOTE: Wash the funnel thoroughly after use.
(3)
Set the FIX-WASH rack into the tank, being careful not to splash any of the solution.
(4)
Fill the DEV tank with solution to just slightly overflow the tank.
(5)
Use the beaker provided to measure out the specified quantity(ex. 78 p) of starter.
(6)
Pour the starter into the DEV tank through the starter fluid inlet.
NOTE :
"Air rock" may occur after finishing this procedure.Be sure to activate the Circulation
pump to check if the solutions circulate normally.
14-14. Checking the Internal Clock
(1) Check the internal clock, and adjust it if its time deviation is unacceptably large.
(refer to "6-5. Time Setup & Display", p.61)
(2) Remove the battery if log information (errors, process count, operation ON/OFF) and internal
clock are not necessary. Other functions and performance of SRX-101A will not be affected
even in operation without the battery.
(refer to "11-5. Replacing Backup Battery", p.101)
INSTALLATION INSTRUCTIONS AND PRECAUTIONS
Full Marks
(in each processing tank)
SERVICE MANUAL