INSTALLATION PROCEDURE
4) Fix the knob plate with the three screws after removing a round plate. (Fig. C)
5) Connect the connector JP18 to the JJ18 on the CPU board and the earth (ground) wire. (Fig. C)
JJ18
CPU Board
Motor control board
Fig. C
Fix this earth to the earth terminal.
6) Adjust the pointer of the knob at the notch of the knob plate. (Fig. D)
7) Assemble the covers onto the main body with the knob pointer fixed at the center (Fig. A).
Pay attention to keep the knob not contracted with the surrounding cover.
8) Stick the dial plate with the knob pointer matching at 34 scale of the dial plate. Pay attention to keep
the knob not contracted with the dial plate. (Fig. E)
Dial Plate
Knob Plate
Notch
34
Pointer
30
38
Fig. D
Fig. E
SERVICE MANUAL