14. Supply pumps (P3, P4) will
not respond to frequency
(DIPSW 1-2) setting
change.
Is DIP SW1-2 in the ON
position?
YES
Replace the CPU board.
15. Initial or replenish
solution supply not
operating properly.
Replace the CPU board.
16. Supply pumps (P3,P4)
will not stop.
Is the film entry detector ON
("READY" lamp OFF)?
NO
Replace the CPU board.
NO
Set DIPSW1-2 to ON.
REMINDER :
In case of 50Hz, "in the OFF position."
NOTE:
Initial supply is the amount of solution supplied immediately after the
[RUN] SW is pressed ON.
Replenish supply is the amount of solution if no film has been processed
over an hour's duration.
REMINDER :
Be aware that initial solution supply may
not be carried out due to the timing of
[RUN] SW ON.
Check if there is any film
YES
jammed at the detector or
not.
REMINDER :
Initial/replenish solution supply may require the supply pumps to
operate for several ten minutes.
TROUBLE SHOOTING
If all of the checks are OK,
replace the film entry
detector unit.
SERVICE MANUAL
39