INSTALLATION INSTRUCTIONS AND PRECAUTIONS
14-13. Filling the Tanks
a. Normal Procedure
(1)
Remove the top cover of the main unit.
(2)
Close all the drainage valves
(3)
Open the facility's water utility valve.
(4)
Plug in the power cord.
The [RUN] SW Lamp on the control panel will light up.
NOTE : Do not press the [RUN] SW.
(5)
Press and hold down the [Repl.] SW (about 5
seconds) until the "READY" Lamp begins flashing.
The tanks will begin to be supplied with processing solution until the "READY" Lamp stops
flashing, indicating that the tanks are full.
This process takes about 22 minutes at 50 Hz and about 18 minutes at 60 Hz.
(6)
Use the beaker(measuring cup) provided to measure out the specified quantity(ex. 78 p) of
starter.
(7)
Pour the starter into the DEV tank through the starter fluid gap shown as below.
(8)
Close the top cover.
Specified quantity
(ex. 78P)
123
SERVICE MANUAL
Measuring Cup
Starter Fluid
Press and hold down
about 5 seconds
Starter Fluid Inlet
RUN
READY
RUN