INSTALLATION PROCEDURE
12) String the control panel lead wire through the hexagonal hole in the switch box(notch facing down)
so that the control panel is protruding out of the hole.
Approx. 20cm of play
13) String the connector end of the lead wire through the hexagonal hole in the film insertion table for
the control panel and the notch provided. Make sure you have about 20cm of play in the wire.
14) Now bring the film insertion table up against the SRX-101A main unit, and string the connector end
of the lead wire through the elongated hole in the main unit until the connector is hanging in front of
the CPU board. At this point, fix the film insertion table to the main unit with the screw provided.
15) Reconnect the lead wire to the CPU board and fix the ground wire terminal with a screw. Make
sure that there is enough play in the wire so that the CPU board is not being pulled or strained.
16) Attach the control panel switch box to a convenient place on the light shield panel.
Note : The control panel may also be attached directly to either the left or right side of the film
insertion table, though it will not be easily seen.
17) Attach Packing(G) to the right and left hand sides of the outer cover of the electrical components
as shown in the diagram below.
139
SERVICE MANUAL
Notch
Insertion Table
Electrical Components Unit Outer Cover
Packing(G)
Hexagonal hole
Lead Wire
Notch
Packing(G)
Control Panel
Switch Box