INSTALLATION INSTRUCTIONS AND PRECAUTIONS
14-7. Water Lines
(1)
Connect a reducer bushing with a packing to the facility's water utility outlet.
(2)
Connect the processor's water supply hose to the reducer bushing.
Facility Water
Utility Outlet
(3)
Acceptable Wash Water Temperature
Required Water Flow .................................................................0.7~1.7R(0.19~0.47gal.)/min
Required Water Pressure ..........................................................29.4~784kpa(0.3~8kgf)/cm
(4)
If water pressure is below 29.4kpa(0.3kgf)/cm
be maintained.
In this case, a malfunction may occur. Increase the water pressure.
(5)
If water pressure is exceeding 784kpa(8kgf)/cm
pressure.
Excessive water pressure applied to the water line may cause the damage of the hoses.
14-8. Setting the Feed Tray in Place
(1)
Use the pan-head screws provided to attach the feed tray to the main unit.
(2)
Control panel can be attached to either side of the main unit.
Select the side which is more convenient to operate, considering the operating environment.
14-9.Cleaning the Racks
(1)
Remove the DEV and FIX racks from the main unit.
(2)
Scrub each rack(including the rollers) with a wet sponge to remove any dirt or grim and then
rinse with plain water. Lean the racks so that the remaining water can evacuate from the
racks.
119
SERVICE MANUAL
Packing
SRX-101A Water Supply Hose
PF3/4
PF1/2
5:~30:(41<~86<)
2
, over 0.7R(0.19gal.)/min Wash water cannot
2
, use flow restrictor to reduce the water
2