INSTALLATION INSTRUCTIONS AND PRECAUTIONS

14. Installation Instructions and Precautions
14-1. Space Requirements
The accompanying diagram shows the minimum space required to operate, maintain and repair the
SRX-101A. Any space smaller than these dimensions will result in longer shutdown times during
maintenance and repair.
14-2. Setting the Main Unit in place (Refer to the Diagram Below)
(1)
Place the Main Unit on the Stand (Option), a desk or a work bench so that the supply and
drainage valves are easily accessible.
(2)
The SRX-101A, when filled with processing solutions, weighs approximately 47 kg. Make sure that
the support to be used is more than capable of bearing this weight and also resistant to the
corrosive effects of the developer (PH 11) and fixer (PH 4) chemicals.
(3)
Do not install the processor in high vibration areas or where it cannot be properly leveled.
(4)
Place the SRX-101A at a height from the floor that fulfills both of the following conditions.
a. The liquid surface inside the replenisher tank is below the full marks of the main unit
processing tanks.
b. The top of the main unit support is less than 130cm from the floor or other surface where the
replenish tank is standing.
Max.130cm
115
SERVICE MANUAL
At least 60cm
SRX–101A
At
least
25cm
At least 60cm
Drainage Valves
Adjustable Stands
At least
60cm
Binding Tape
with thread
Developer Supply
Hose (Yellow)
Fixer Supply
Hose (Red)
Water Supply Hose
Developer Drainage
Hose (Yellow)
Fixer Drainage
Hose (Red)
Wash Water Drainage Hose
(Thick, Transparent)
Overflow Drainage Hose
(Thin, Transparent)