8.9 Checking Sieve Bed Pressure (Model IRC5PO2VAW Only)
DANGER!
– To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet before servicing.
Disassembly
Product Tank with Regulator Manufactured Before 7/1/2012
1.
Turn power Off (O) and unplug unit.
2.
Remove cabinet. Refer to 6.1 Removing Cabinet, page 17.
3.
Turn unit so the back faces you.
Remove tie wrap and disengage pressure sensor tubing A from product tank.
4.
Installation
1.
Insert the 1/8'' "T" fitting into the ends of the tubing.
2.
Secure tie wraps to each end of the tubing 1/8'' "T" fitting connection.
3.
Attach the 1/8'' x 2-inch tubing to the new 1/8'' "T" fitting and secure with tie wrap.
4.
Insert plug into the open end of the1/8'' x 2-inch tubing.
5.
Secure with tie wrap.
1154245-F-02
Product Tank with Regulator Manufactured After 7/1/2012
A
Testing
47