Invacare® Perfecto₂™ Oxygen Concentrator
6.2 Replacing Compressor Assembly
6.2.1 Replacing Compressor Assembly Before September 2015
DANGER!
Risk of Death, Injury, or Damage
To prevent electrical shock
– ALWAYS disconnect concentrator from electrical outlet before servicing.
Refer to recycling information following local governing ordinances and recycling plans regarding disposal of the concentrator or
components.
PE valve not shown.
PE valve not shown.
1.
Turn power Off (O) and unplug the unit.
2.
Remove the cabinet. Refer to 6.1 Removing Cabinet, page 17.
3.
Cut the tie wrap A that secures the intake hose B to the compressor assembly C.
4.
Disconnect intake hose from compressor assembly.
5.
Remove compressor wires from wire clamps or tie-wraps (not shown).
6.
Disconnect compressor connector from the main harness connector (not shown).
Disconnect capacitor wires D from the top of the capacitor E.
7.
18
IRC5PO2AW
PE valve not shown.
IRC5PO2VAW
Filter and PE valve not shown.
1154245-F-02