Invacare® Perfecto₂™ Oxygen Concentrator
6.3.2 Replacing Capacitor After September 2015
IRC5PO2AW
Filter and PE valve not shown.
1.
Turn power Off (O) and unplug the unit.
2.
Remove the cabinet. Refer to 6.1 Removing Cabinet, page 17.
Using needlenose pliers, disconnect the spade connectors A from the capacitor terminals B .
3.
Cut the tie wraps (not shown)that secures the capacitor C to the vertical sound box and remove the capacitor.
4.
5.
Attach non-skid pad (1072004) to capacitor.
6.
Install the new capacitor by reversing STEPS 3 and 4.
7.
Run unit and inspect for leaks. Refer to 8.1 Leak Test, page 41.
8.
If no leaks detected, turn power Off (O) and unplug the unit.
9.
Re-install the cabinet. Refer to 6.1 Removing Cabinet, page 17 and reverse all steps.
6.4 Replacing P. E. Valve
DANGER!
Risk of Death, Injury, or Damage
To prevent electrical shock
– ALWAYS disconnect concentrator from electrical outlet before servicing.
22
IRC5PO2VAW
Filter and PE valve not shown.
1154245-F-02