Replacing Capacitor - Invacare Platinum IRC5LXO2AW Service Manual

Hide thumbs Also See for Platinum IRC5LXO2AW:
Table of Contents

Advertisement

Clamp
Intake
Hose
Mounting
Screws
FIGURE 7.2 Replacing Compressor Assembly - 9 Liter Models

Replacing Capacitor

1. Turn power Off (O) and unplug the unit.
2. Remove the cabinet. Refer to Removing Cabinet on page 25.
3. Remove protective boot to reveal capacitor wires.
4. Disconnect the spade connectors from the capacitor terminals (not shown).
5. Remove the screw that secures the capacitor clamp to the soundbox brace.
6. Remove the capacitor from the concentrator base assembly.
7. Install new capacitor by reversing STEPS 3-6.
8. Run unit and inspect for leaks. Refer to Leak Test on page 71.
9. If no leaks detected, turn power Off (O) and unplug the unit.
10. Reinstall cabinet. Refer to Removing Cabinet on page 25.
Protective Boot
Part No 1118396
FIGURE 7.3 Replacing Capacitor
37
SECTION 7—COMPRESSOR
Capacitor Clamp
Capacitor
Screw
Soundbox Brace
Motor/Main
Harness
Connection
Compressor
Output Hose
Protective Boot
Capacitor
Compressor
Restraint Bracket
®
Platinum
Oxygen