TiX500, TiX520, TiX560, TiX580
Users Manual
Surfaces with an emissivity of
consistent actual temperatures. The lower the emissivity is the greater the
potential of error is when the Imager calculates the temperature measurement
because more of the energy reaching the camera is specified as background
temperature. This is also true even when adjustments to the emissivity and
reflected background adjustments are performed properly.

Spot Markers

Use fixed-temperature spot markers to highlight a region on the display before you save an
image.
To set a marker:
1. Select Measurement > Markers.
2. Select an option.
3. Push  to set the marker option and go to the Move Marker display.
The Move Marker icon shows on the display and the labels on the function buttons change
to Done, Next, and Cancel.
To change the marker position on the display:
 
1. Push
 
2. Push  to select the next marker.
3. Repeat for the rest of the markers.
4. Push  when done.

Spot Box

Use the Spot Box feature to adjust a temperature measurement zone (box) to center on the
target. This zone expands and contracts to different levels within the infrared image. The zone
shows an approximate maximum (MAX), average (AVG), and minimum (MIN) temperature
measurement in that area.
When using the Spot Box, the level and span of the Imager adjusts to the thermal
scene within the Spot Box.
18
Shop for Fluke products online at:
Note
<
0.60 make it difficult to determine reliable and
to move the marker location on the image.
Note
MyFlukeStore
www.
1.888.610.7664
.com