Figure 3-4. Changing the Contrast Setting

Digital Modality with Remote Keypad

1. On the modality, select and display the image you want to use for printing
contrast samples.
2. On the DryView 8600 remote keypad, select a format other than 1. (For
example, use 2, 4, or 6.)
3. Press Acquire on the keypad.
4. On the local panel, press
host-controlled modality from which you will print contrast samples.
5. On the local panel, press
then press
. (The DryView 8600 automatically prints 12 copies of the
Enter
previously acquired image at different contrasts.)
6. Select the image that you prefer, and note the image number associated with it.
7. On the local panel, press
8. Following the instructions on the display, change the contrast setting to the
number previously selected from the contrast samples, then press
9. After setting the contrast, press
2000 March Rev. B
8186546
B
A
or
to select the menu that represents the
Setup
Setup
to go to item 3, Print Contrast Samples, and
to go to item 2, Contrast.
to return the DryView 8600 to service.