Table 2-4 2.4-mm (50 Ω) Device Connector Type
Network Analyzer: HP 8722ET Standard or Option 004 Attenuator
Calibration Kit: HP 85056A (2.4-mm, 50 Ω)
Cables: HP 85133E Cables
Calibration: One-Port, Response, or Enhanced Response
IF BW = 10 Hz, Avg = off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Reflection Measurements
Directivity (dB)
Source Match (dB)
Load Match (dB)
One-Port Cal
Tracking
Magnitude (dB)
Phase (deg)
Transmission Measurements
Source Match (dB)
Enhanced
Response Cal
Response Only
Cal
Load Match (dB)
Tracking
Enhanced
Response Cal
Magnitude (dB)
Phase (deg)
Response Only
Cal
Magnitude (dB)
Phase (deg)
Uncertainty graphs are on the following page.
a. One-port or enhanced response calibration.
b. Enhanced response or response only calibration.
0.05 to 2 GHz
2 to 8 GHz
a
42
41
22
±(0.005 + .03/°C)
±(0.010 + .03/°C)
±(0.033 + .1/°C)
±(0.066 + .3/°C)
b
41
16
22
±(0.018 + .03/°C)
±(0.026 + .03/ °C)
±(0.119 + .1/°C)
±(0.172 + .3/°C)
±(0.109 + .03/°C)
±(0.172 + .03/°C)
±(0.719 + .01/°C)
±(1.135 + .3/°C)
HP 8719/20/22ET Specifications and Characteristics
Corrected System Performance (HP 8722ET)
Specification
8 to 20 GHz
42
42
38
38
20
20
±(0.010 + .04/°C)
±(0.066 + .5/°C)
38
38
14
11
20
20
±(0.031 + .04/°C)
±(0.205 + .5/°C)
±(0.241 + .04/°C)
±(1.59 + .5/°C)
20 to 40 GHz
38
33
15
±(0.021 + .06/ °C)
±(0.139 + 1.0/°C)
33
10
15
±(0.069 + .06/°C)
±(0.455 + 1.0/°C)
±(0.475 + .06/°C)
±(3.135 + 1.0/°C)
2-7